GHS-02 Five Point Gradient Film Heat Sealer

Application

Gradient Heat Seal Tester is professionally designed for the precise measurement of five groups of sealing temperature, time and pressure. How well the material heat seals is expressed quantitatively as seal strength as a function of the sealing conditions of temperature, time and pressure.

Features
  • PLC control and 7 inch touching screen operation
  • Five groups sealing temperature can be measured with one test
  • Proximity and pressure sensor for precise dwell time and sealing pressure control
  • P.I.D. temperature control
  • Foot petal equipped
  • Uniform heating by aluminum encapsulated sealing jaws
  • Independent heating of upper and lower sealing jaws
  • Sealing jaw customization available
  • RS 232 port with professional software (optional)

Key Data

Sealing Temp.                              Room Temp.~300℃

Temp. Gradient                            ≤20℃

Deviation                                        ±0.2℃

Sealing Time                                  0.1s~9999s

Sealing Pressure                           0.15 MPa~0.7 MPa

Seal Jaws                                        40mm×10mm×5

Gas Pressure                                  0.7 MPa

Port Size                                           Ф6 mm PU Hose

Power                                               AC 220V 50Hz