{"id":7924,"date":"2026-04-21T14:25:58","date_gmt":"2026-04-21T06:25:58","guid":{"rendered":"https:\/\/www.qualitester.com\/?page_id=7924"},"modified":"2026-04-21T14:25:58","modified_gmt":"2026-04-21T06:25:58","slug":"iso-9187","status":"publish","type":"page","link":"https:\/\/www.qualitester.com\/sk\/standards\/iso-9187\/","title":{"rendered":"ISO 9187"},"content":{"rendered":"<h2 data-section-id=\"1qhubs7\" data-start=\"280\" data-end=\"337\">Pochopenie normy ISO 9187 a jej \u00falohy v kvalite ampuliek<\/h2>\n<p data-start=\"339\" data-end=\"727\"><strong data-start=\"339\" data-end=\"351\">ISO 9187<\/strong> definuje technick\u00e9 po\u017eiadavky a sk\u00fa\u0161obn\u00e9 met\u00f3dy pre sklenen\u00e9 ampulky pou\u017e\u00edvan\u00e9 v injek\u010dn\u00fdch liekoch. Zabezpe\u010duje, aby ampulky sp\u013a\u0148ali pr\u00edsne krit\u00e9ri\u00e1 pre <strong data-start=\"512\" data-end=\"579\">chemick\u00e1 odolnos\u0165, mechanick\u00e1 pevnos\u0165 a bezpe\u010dnos\u0165 po\u010das pou\u017e\u00edvania<\/strong>. Pre v\u00fdrobcov liekov a laborat\u00f3ri\u00e1 kontroly kvality m\u00e1 s\u00falad s normou ISO 9187 priamy vplyv na integritu produktu a bezpe\u010dnos\u0165 pacienta.<\/p>\n<p data-start=\"729\" data-end=\"769\">Norma sa sklad\u00e1 z dvoch hlavn\u00fdch \u010dast\u00ed:<\/p>\n<ul data-start=\"770\" data-end=\"894\">\n<li data-section-id=\"770e0x\" data-start=\"770\" data-end=\"821\"><strong data-start=\"772\" data-end=\"786\">ISO 9187-1<\/strong>: V\u0161eobecn\u00e9 po\u017eiadavky na ampulky<\/li>\n<li data-section-id=\"igh88z\" data-start=\"822\" data-end=\"894\"><strong data-start=\"824\" data-end=\"838\">ISO 9187-2<\/strong>: Osobitn\u00e9 po\u017eiadavky na ampulky s jednobodov\u00fdm rezom (OPC)<\/li>\n<\/ul>\n<p data-start=\"896\" data-end=\"1081\">Tieto dokumenty stanovuj\u00fa jednotn\u00e9 postupy testovania vr\u00e1tane <strong data-start=\"960\" data-end=\"985\">hydrolytick\u00e1 odolnos\u0165<\/strong> a <strong data-start=\"990\" data-end=\"1021\">sk\u00fa\u0161ka pevnosti ampuliek pri pretrhnut\u00ed<\/strong>, ktor\u00e9 tvoria z\u00e1klad syst\u00e9mov zabezpe\u010denia kvality ampuliek.<\/p>\n<h2 data-section-id=\"1yunlpw\" data-start=\"1088\" data-end=\"1140\">ISO 9187-1: Po\u017eiadavky a z\u00e1klady testovania<\/h2>\n<p data-start=\"1142\" data-end=\"1242\">Stanovuje z\u00e1kladn\u00e9 krit\u00e9ri\u00e1 v\u00fdkonu pre ampulky so zameran\u00edm na tri k\u013e\u00fa\u010dov\u00e9 aspekty:<\/p>\n<h3 data-section-id=\"1dy2xvd\" data-start=\"1244\" data-end=\"1283\">Po\u017eiadavka na hydrolytick\u00fa odolnos\u0165<\/h3>\n<p data-start=\"1284\" data-end=\"1531\">Ampulky musia vykazova\u0165 <strong data-start=\"1310\" data-end=\"1345\">vysok\u00e1 odolnos\u0165 proti p\u00f4sobeniu vody<\/strong> na ich vn\u00fatornom povrchu. Pod\u013ea normy musia by\u0165 v\u00fdsledky v s\u00falade s <strong data-start=\"1425\" data-end=\"1456\">Klasifik\u00e1cia ISO 4802 HC1<\/strong>, \u010d\u00edm sa zabezpe\u010d\u00ed minim\u00e1lne vyl\u00fahovanie alkalick\u00fdch zlo\u017eiek do injek\u010dn\u00fdch roztokov.<\/p>\n<h3 data-section-id=\"vik92a\" data-start=\"1533\" data-end=\"1556\">Kvalita \u017e\u00edhania<\/h3>\n<p data-start=\"1557\" data-end=\"1740\">Spr\u00e1vne \u017e\u00edhanie zni\u017euje vn\u00fatorn\u00e9 nap\u00e4tie v skle. \u0160tandardn\u00e9 limity <strong data-start=\"1628\" data-end=\"1664\">optick\u00e9 spomalenie na \u226450 nm\/mm<\/strong>, \u010d\u00edm sa zabezpe\u010d\u00ed stabilita kon\u0161trukcie a zn\u00ed\u017ei sa riziko zlomenia pri manipul\u00e1cii.<\/p>\n<h3 data-section-id=\"ebplwc\" data-start=\"1742\" data-end=\"1774\">V\u00fdkonnos\u0165 sily pri prelomen\u00ed<\/h3>\n<p data-start=\"1775\" data-end=\"1889\">Str\u00e1nka <strong data-start=\"1779\" data-end=\"1810\">sk\u00fa\u0161ka pevnosti ampuliek pri pretrhnut\u00ed<\/strong> vyhodnocuje silu potrebn\u00fa na oddelenie krku od tela. T\u00fdm sa zabezpe\u010d\u00ed:<\/p>\n<ul data-start=\"1890\" data-end=\"1974\">\n<li data-section-id=\"p1tnbf\" data-start=\"1890\" data-end=\"1918\">Riaden\u00e1 sila otv\u00e1rania<\/li>\n<li data-section-id=\"k8llgo\" data-start=\"1919\" data-end=\"1943\">\u010cist\u00e1 a bezpe\u010dn\u00e1 prest\u00e1vka<\/li>\n<li data-section-id=\"1i25w4v\" data-start=\"1944\" data-end=\"1974\">Zn\u00ed\u017een\u00e9 riziko kontamin\u00e1cie<\/li>\n<\/ul>\n<p data-start=\"1976\" data-end=\"2024\">Testovanie si vy\u017eaduje \u0165ahov\u00fd testovac\u00ed stroj s:<\/p>\n<ul data-start=\"2025\" data-end=\"2078\">\n<li data-section-id=\"1li5qup\" data-start=\"2025\" data-end=\"2047\"><strong data-start=\"2027\" data-end=\"2047\">R\u00fdchlos\u0165: 10 mm\/min<\/strong><\/li>\n<li data-section-id=\"1gqcesy\" data-start=\"2048\" data-end=\"2078\"><strong data-start=\"2050\" data-end=\"2078\">Rozsah sily: do 200 N<\/strong><\/li>\n<\/ul>\n<h2 data-section-id=\"1oiwwvg\" data-start=\"2085\" data-end=\"2140\">ISO 9187-2: Ampulky OPC a \u010fal\u0161ie po\u017eiadavky<\/h2>\n<p data-start=\"2142\" data-end=\"2236\">Roz\u0161iruje po\u017eiadavky na ampulky OPC, ktor\u00e9 sa \u0161iroko pou\u017e\u00edvaj\u00fa na bezpe\u010dnej\u0161ie manu\u00e1lne otv\u00e1ranie.<\/p>\n<h3 data-section-id=\"1s0v4j0\" data-start=\"2238\" data-end=\"2256\">K\u013e\u00fa\u010dov\u00e9 pr\u00edrastky:<\/h3>\n<ul data-start=\"2257\" data-end=\"2500\">\n<li data-section-id=\"1llhxxn\" data-start=\"2257\" data-end=\"2335\"><strong data-start=\"2259\" data-end=\"2289\">Umiestnenie bodu zlomu<\/strong>: Mus\u00ed zosta\u0165 v rozmedz\u00ed \u00b11 mm od osi<\/li>\n<li data-section-id=\"120evac\" data-start=\"2336\" data-end=\"2417\"><strong data-start=\"2338\" data-end=\"2359\">Tepeln\u00e1 stabilita<\/strong>: Mus\u00ed vydr\u017ea\u0165 zahriatie na 120 \u00b0C s n\u00e1sledn\u00fdm ochladen\u00edm vodou<\/li>\n<li data-section-id=\"1uhlasw\" data-start=\"2418\" data-end=\"2500\"><strong data-start=\"2420\" data-end=\"2448\">Odolnos\u0165 vo\u010di steriliz\u00e1cii<\/strong>: Mus\u00ed vydr\u017ea\u0165 \u010distiace a steriliza\u010dn\u00e9 procesy<\/li>\n<\/ul>\n<p data-start=\"2502\" data-end=\"2604\">Tieto krit\u00e9ri\u00e1 zabezpe\u010duj\u00fa konzistentn\u00fa pou\u017eite\u013enos\u0165 a spo\u013eahlivos\u0165 v re\u00e1lnych farmaceutick\u00fdch podmienkach.<\/p>\n<h2 data-section-id=\"102k1ie\" data-start=\"2611\" data-end=\"2641\">Sk\u00fa\u0161ka pevnosti pri pretrhnut\u00ed ampulky<\/h2>\n<p data-start=\"2643\" data-end=\"2730\">Str\u00e1nka <strong data-start=\"2647\" data-end=\"2678\">sk\u00fa\u0161ka pevnosti ampuliek pri pretrhnut\u00ed<\/strong> je kritick\u00e9 mechanick\u00e9 hodnotenie pod\u013ea normy ISO 9187.<\/p>\n<h3 data-section-id=\"oz1o92\" data-start=\"2732\" data-end=\"2752\">Princ\u00edp testu<\/h3>\n<p data-start=\"2753\" data-end=\"2934\">Na ampulku sa p\u00f4sob\u00ed riadenou silou kolmo (90\u00b0) na jej os, a\u017e k\u00fdm ned\u00f4jde k jej roztrhnutiu. Syst\u00e9m zaznamen\u00e1va <strong data-start=\"2868\" data-end=\"2892\">hodnota sily pri pretrhnut\u00ed<\/strong>, ktor\u00e9 musia by\u0165 v r\u00e1mci stanoven\u00fdch limitov.<\/p>\n<h3 data-section-id=\"1lrggvp\" data-start=\"2936\" data-end=\"2954\">Pre\u010do je to d\u00f4le\u017eit\u00e9<\/h3>\n<ul data-start=\"2955\" data-end=\"3126\">\n<li data-section-id=\"1h9xoo4\" data-start=\"2955\" data-end=\"3015\">Zabezpe\u010duje <strong data-start=\"2965\" data-end=\"3013\">jednoduch\u00e9 otv\u00e1ranie pre zdravotn\u00edckych pracovn\u00edkov<\/strong><\/li>\n<li data-section-id=\"wkov47\" data-start=\"3016\" data-end=\"3070\">Zabra\u0148uje <strong data-start=\"3027\" data-end=\"3068\">nadmern\u00e1 sila, ktor\u00e1 m\u00f4\u017ee sp\u00f4sobi\u0165 zranenie<\/strong><\/li>\n<li data-section-id=\"g9ynyp\" data-start=\"3071\" data-end=\"3126\">Z\u00e1ruky <strong data-start=\"3084\" data-end=\"3126\">\u010dist\u00fd lom bez \u00falomkov skla<\/strong><\/li>\n<\/ul>\n<h3 data-section-id=\"1blat2i\" data-start=\"3128\" data-end=\"3154\">Odpor\u00fa\u010dan\u00e9 rie\u0161enie<\/h3>\n<p><a href=\"https:\/\/www.qualitester.com\/sk\/produkt-2\/how-to-test-the-force-about-ampoule-breaking\/\"><img fetchpriority=\"high\" decoding=\"async\" class=\"alignleft wp-image-7925 size-medium\" src=\"https:\/\/www.qualitester.com\/wp-content\/uploads\/2026\/04\/ISO-9187-Ampoule-Break-Force-Tester-300x300.webp\" alt=\"Tester sily pretrhnutia ampuliek ISO 9187\" width=\"300\" height=\"300\" srcset=\"https:\/\/www.qualitester.com\/wp-content\/uploads\/2026\/04\/ISO-9187-Ampoule-Break-Force-Tester-300x300.webp 300w, https:\/\/www.qualitester.com\/wp-content\/uploads\/2026\/04\/ISO-9187-Ampoule-Break-Force-Tester-150x150.webp 150w, https:\/\/www.qualitester.com\/wp-content\/uploads\/2026\/04\/ISO-9187-Ampoule-Break-Force-Tester-12x12.webp 12w, https:\/\/www.qualitester.com\/wp-content\/uploads\/2026\/04\/ISO-9187-Ampoule-Break-Force-Tester-100x100.webp 100w, https:\/\/www.qualitester.com\/wp-content\/uploads\/2026\/04\/ISO-9187-Ampoule-Break-Force-Tester.webp 600w\" sizes=\"(max-width: 300px) 100vw, 300px\" \/><\/a><\/p>\n<p data-start=\"3155\" data-end=\"3508\">Str\u00e1nka <strong data-start=\"3159\" data-end=\"3193\">BST-01 <a href=\"https:\/\/www.qualitester.com\/sk\/produkt-2\/how-to-test-the-force-about-ampoule-breaking\/\">Tester sily pretrhnutia ampulky<\/a><\/strong> poskytuje vysoko opakovate\u013en\u00fd a automatizovan\u00fd pr\u00edstup k testovaniu sklenen\u00fdch ampuliek. Jeho <strong data-start=\"3271\" data-end=\"3345\">presn\u00e1 regul\u00e1cia ot\u00e1\u010dok, vymenite\u013en\u00e9 pr\u00edpravky a bezpe\u010dnostn\u00e1 ochrana<\/strong> je ide\u00e1lny pre laborat\u00f3ri\u00e1 kontroly kvality aj pre prostredie v\u00fdskumu a v\u00fdvoja. Syst\u00e9m v\u00fdrazne zvy\u0161uje efektivitu testovania pri zachovan\u00ed s\u00faladu so \u0161tandardn\u00fdmi po\u017eiadavkami.<\/p>\n<h2 data-section-id=\"1utx21z\" data-start=\"3515\" data-end=\"3556\">Sk\u00fa\u0161ka hydrolytickej odolnosti ampuliek<\/h2>\n<p data-start=\"3558\" data-end=\"3669\">Str\u00e1nka <strong data-start=\"3562\" data-end=\"3604\">sk\u00fa\u0161ka hydrolytickej odolnosti ampuliek<\/strong> hodnot\u00ed chemick\u00fa odolnos\u0165 skla pri p\u00f4soben\u00ed vody.<\/p>\n<h3 data-section-id=\"oz1o92\" data-start=\"3671\" data-end=\"3691\">Princ\u00edp testu<\/h3>\n<p data-start=\"3692\" data-end=\"3855\">Sklo sa rozdrv\u00ed na ur\u010dit\u00fa ve\u013ekos\u0165 \u010dast\u00edc a za kontrolovan\u00fdch podmienok sa vystav\u00ed p\u00f4sobeniu vody. Na ur\u010denie odolnosti sa meria mno\u017estvo uvo\u013enenej z\u00e1sady.<\/p>\n<h3 data-section-id=\"113ldup\" data-start=\"3857\" data-end=\"3885\">K\u013e\u00fa\u010dov\u00fd \u0161tandardn\u00fd odkaz<\/h3>\n<p data-start=\"3886\" data-end=\"3986\">Testovanie sa vykon\u00e1va pod\u013ea met\u00f3d ISO 4802, ako to vy\u017eaduje norma ISO 9187, \u010d\u00edm sa zabezpe\u010duje konzistentnos\u0165 v laborat\u00f3ri\u00e1ch.<\/p>\n<h3 data-section-id=\"1lrggvp\" data-start=\"3988\" data-end=\"4006\">Pre\u010do je to d\u00f4le\u017eit\u00e9<\/h3>\n<ul data-start=\"4007\" data-end=\"4168\">\n<li data-section-id=\"r2xdzf\" data-start=\"4007\" data-end=\"4066\">Zabra\u0148uje <strong data-start=\"4018\" data-end=\"4064\">chemick\u00e9 interakcie s injek\u010dn\u00fdmi liekmi<\/strong><\/li>\n<li data-section-id=\"novl3u\" data-start=\"4067\" data-end=\"4114\">Udr\u017eiava <strong data-start=\"4079\" data-end=\"4112\">stabilita a trvanlivos\u0165 lieku<\/strong><\/li>\n<li data-section-id=\"am3akf\" data-start=\"4115\" data-end=\"4168\">Zabezpe\u010duje <strong data-start=\"4125\" data-end=\"4168\">dodr\u017eiavanie pr\u00e1vnych predpisov na glob\u00e1lnych trhoch<\/strong><\/li>\n<\/ul>\n<h3 data-section-id=\"1blat2i\" data-start=\"4170\" data-end=\"4196\">Odpor\u00fa\u010dan\u00e9 rie\u0161enie<\/h3>\n<p><a href=\"https:\/\/www.qualitester.com\/sk\/produkt-2\/sale-glass-grains-sampler-for-hydrolytic-resistance\/\"><img decoding=\"async\" class=\"alignleft wp-image-7926 size-medium\" src=\"https:\/\/www.qualitester.com\/wp-content\/uploads\/2026\/04\/ISO-9187-Glass-Grain-Hydrolytic-Resistance-Sampler-300x300.webp\" alt=\"ISO 9187 Vzorkova\u010d na hydrolytick\u00fa odolnos\u0165 sklenen\u00fdch z\u0155n\" width=\"300\" height=\"300\" srcset=\"https:\/\/www.qualitester.com\/wp-content\/uploads\/2026\/04\/ISO-9187-Glass-Grain-Hydrolytic-Resistance-Sampler-300x300.webp 300w, https:\/\/www.qualitester.com\/wp-content\/uploads\/2026\/04\/ISO-9187-Glass-Grain-Hydrolytic-Resistance-Sampler-150x150.webp 150w, https:\/\/www.qualitester.com\/wp-content\/uploads\/2026\/04\/ISO-9187-Glass-Grain-Hydrolytic-Resistance-Sampler-12x12.webp 12w, https:\/\/www.qualitester.com\/wp-content\/uploads\/2026\/04\/ISO-9187-Glass-Grain-Hydrolytic-Resistance-Sampler-100x100.webp 100w, https:\/\/www.qualitester.com\/wp-content\/uploads\/2026\/04\/ISO-9187-Glass-Grain-Hydrolytic-Resistance-Sampler.webp 600w\" sizes=\"(max-width: 300px) 100vw, 300px\" \/><\/a><\/p>\n<p data-start=\"4197\" data-end=\"4305\">Str\u00e1nka <strong data-start=\"4201\" data-end=\"4253\">GHR-01A <a href=\"https:\/\/www.qualitester.com\/sk\/produkt-2\/sale-glass-grains-sampler-for-hydrolytic-resistance\/\">Tester hydrolytickej odolnosti sklenen\u00fdch z\u0155n<\/a><\/strong> automatizuje pr\u00edpravu a preosievanie vzoriek, \u010d\u00edm zabezpe\u010duje:<\/p>\n<ul data-start=\"4306\" data-end=\"4421\">\n<li data-section-id=\"1xas01l\" data-start=\"4306\" data-end=\"4348\"><strong data-start=\"4308\" data-end=\"4346\">D\u00f4sledn\u00fd v\u00fdber ve\u013ekosti \u010dast\u00edc<\/strong><\/li>\n<li data-section-id=\"7hw5ca\" data-start=\"4349\" data-end=\"4381\"><strong data-start=\"4351\" data-end=\"4379\">Zv\u00fd\u0161en\u00e1 bezpe\u010dnos\u0165 obsluhy<\/strong><\/li>\n<li data-section-id=\"wyf3le\" data-start=\"4382\" data-end=\"4421\"><strong data-start=\"4384\" data-end=\"4421\">Vy\u0161\u0161ia opakovate\u013enos\u0165 a presnos\u0165<\/strong><\/li>\n<\/ul>\n<p data-start=\"6367\" data-end=\"6790\">\n<h2 data-section-id=\"1qsfy1n\" data-start=\"4972\" data-end=\"5008\">Z\u00edskajte rie\u0161enie<\/h2>\n<p>Zabezpe\u010dte, aby va\u0161e ampulky sp\u013a\u0148ali po\u017eiadavky normy ISO 9187 pomocou presn\u00e9ho testovania hydrolytickej odolnosti a sily pretrhnutia. Prijat\u00edm \u0161tandardizovan\u00fdch met\u00f3d a automatizovan\u00fdch testovac\u00edch rie\u0161en\u00ed m\u00f4\u017eete zv\u00fd\u0161i\u0165 bezpe\u010dnos\u0165, konzistenciu a zhodu v\u00fdrobkov. Vybavte svoje laborat\u00f3rium spo\u013eahliv\u00fdmi pr\u00edstrojmi na zefekt\u00edvnenie pracovn\u00fdch postupov, zn\u00ed\u017eenie variability a poskytovanie vysokokvalitn\u00fdch farmaceutick\u00fdch obalov, ktor\u00e9 sp\u013a\u0148aj\u00fa glob\u00e1lne regula\u010dn\u00e9 o\u010dak\u00e1vania.<\/p>\n<h2 data-section-id=\"1qsfy1n\" data-start=\"4972\" data-end=\"5008\">\u010casto kladen\u00e9 ot\u00e1zky<\/h2>\n<p data-start=\"5010\" data-end=\"5155\"><strong data-start=\"5010\" data-end=\"5054\">1. Ak\u00fd je hlavn\u00fd \u00fa\u010del normy ISO 9187?<\/strong><br data-start=\"5054\" data-end=\"5057\" \/>Definuje po\u017eiadavky na kvalitu a sk\u00fa\u0161anie sklenen\u00fdch ampuliek pou\u017e\u00edvan\u00fdch v injek\u010dn\u00fdch liekoch.<\/p>\n<p data-start=\"5157\" data-end=\"5310\"><strong data-start=\"5157\" data-end=\"5214\">2. Ak\u00e1 je pr\u00edpustn\u00e1 sila pri pretrhnut\u00ed pod\u013ea normy ISO 9187?<\/strong><br data-start=\"5214\" data-end=\"5217\" \/>Prijate\u013en\u00fd rozsah z\u00e1vis\u00ed od ve\u013ekosti a kon\u0161trukcie ampuliek, ako je uveden\u00e9 v \u0161tandardn\u00fdch tabu\u013ek\u00e1ch.<\/p>\n<p data-start=\"5312\" data-end=\"5456\"><strong data-start=\"5312\" data-end=\"5370\">3. Pre\u010do je hydrolytick\u00e1 odolnos\u0165 pre ampulky rozhoduj\u00faca?<\/strong><br data-start=\"5370\" data-end=\"5373\" \/>Zabra\u0148uje chemickej kontamin\u00e1cii injek\u010dn\u00fdch liekov sp\u00f4sobenej interakciou so sklom.<\/p>\n<p data-start=\"5458\" data-end=\"5604\"><strong data-start=\"5458\" data-end=\"5514\">4. Ako sa vykon\u00e1va sk\u00fa\u0161ka pevnosti pri pretrhnut\u00ed ampulky?<\/strong><br data-start=\"5514\" data-end=\"5517\" \/>Na \u0165ahovom testovacom stroji p\u00f4sob\u00ed sila kontrolovanou r\u00fdchlos\u0165ou, a\u017e k\u00fdm sa ampulka nerozbije.<\/p>\n<p data-start=\"5606\" data-end=\"5779\"><strong data-start=\"5606\" data-end=\"5670\">5. Ak\u00fd je rozdiel medzi normami ISO 9187-1 a ISO 9187-2?<\/strong><br data-start=\"5670\" data-end=\"5673\" \/><a href=\"https:\/\/www.iso.org\/standard\/55920.html\" target=\"_blank\" rel=\"noopener nofollow\">ISO 9187-1<\/a> sa vz\u0165ahuje na v\u0161eobecn\u00e9 ampulky, zatia\u013e \u010do <a href=\"https:\/\/www.iso.org\/standard\/55921.html\" target=\"_blank\" rel=\"noopener nofollow\">ISO 9187-2<\/a> sa zameriava na ampulky OPC s \u010fal\u0161\u00edmi po\u017eiadavkami.<\/p>","protected":false},"excerpt":{"rendered":"<p>Porozumenie norme ISO 9187 a jej \u00falohe v kvalite ampuliek Norma ISO 9187 definuje technick\u00e9 po\u017eiadavky a sk\u00fa\u0161obn\u00e9 met\u00f3dy pre sklenen\u00e9 ampulky pou\u017e\u00edvan\u00e9 v injek\u010dn\u00fdch liekoch. Zabezpe\u010duje, aby ampulky sp\u013a\u0148ali pr\u00edsne krit\u00e9ri\u00e1 t\u00fdkaj\u00face sa chemickej odolnosti, mechanickej pevnosti a bezpe\u010dnosti pri pou\u017e\u00edvan\u00ed. Pre v\u00fdrobcov liekov a laborat\u00f3ri\u00e1 kontroly kvality m\u00e1 dodr\u017eiavanie normy ISO 9187 priamy vplyv na produkt \u2026<\/p>\n<p class=\"read-more\"> <a class=\"\" href=\"https:\/\/www.qualitester.com\/sk\/standards\/iso-9187\/\"> <span class=\"screen-reader-text\">ISO 9187<\/span> \u010c\u00edtajte viac \"<\/a><\/p>","protected":false},"author":4,"featured_media":0,"parent":4948,"menu_order":0,"comment_status":"closed","ping_status":"closed","template":"","meta":{"site-sidebar-layout":"default","site-content-layout":"default","ast-global-header-display":"","ast-main-header-display":"","ast-hfb-above-header-display":"","ast-hfb-below-header-display":"","ast-hfb-mobile-header-display":"","site-post-title":"","ast-breadcrumbs-content":"","ast-featured-img":"","footer-sml-layout":"","theme-transparent-header-meta":"default","adv-header-id-meta":"","stick-header-meta":"","header-above-stick-meta":"","header-main-stick-meta":"","header-below-stick-meta":"","footnotes":""},"class_list":["post-7924","page","type-page","status-publish","hentry"],"_links":{"self":[{"href":"https:\/\/www.qualitester.com\/sk\/wp-json\/wp\/v2\/pages\/7924","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/www.qualitester.com\/sk\/wp-json\/wp\/v2\/pages"}],"about":[{"href":"https:\/\/www.qualitester.com\/sk\/wp-json\/wp\/v2\/types\/page"}],"author":[{"embeddable":true,"href":"https:\/\/www.qualitester.com\/sk\/wp-json\/wp\/v2\/users\/4"}],"replies":[{"embeddable":true,"href":"https:\/\/www.qualitester.com\/sk\/wp-json\/wp\/v2\/comments?post=7924"}],"version-history":[{"count":0,"href":"https:\/\/www.qualitester.com\/sk\/wp-json\/wp\/v2\/pages\/7924\/revisions"}],"up":[{"embeddable":true,"href":"https:\/\/www.qualitester.com\/sk\/wp-json\/wp\/v2\/pages\/4948"}],"wp:attachment":[{"href":"https:\/\/www.qualitester.com\/sk\/wp-json\/wp\/v2\/media?parent=7924"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}