{"id":7567,"date":"2025-11-13T09:51:57","date_gmt":"2025-11-13T01:51:57","guid":{"rendered":"https:\/\/www.qualitester.com\/?page_id=7567"},"modified":"2025-11-13T10:08:10","modified_gmt":"2025-11-13T02:08:10","slug":"iso-7864","status":"publish","type":"page","link":"https:\/\/www.qualitester.com\/nl\/standards\/iso-7864\/","title":{"rendered":"ISO 7864"},"content":{"rendered":"<h1 style=\"text-align: center;\">ISO 7864<br \/>\n<small>Sterile Hypodermic Needle Testing<br \/>\nPenetration and Drag Force &amp; Needle Bonding Strength<\/small><\/h1>\n<p data-start=\"437\" data-end=\"871\"><strong data-start=\"456\" data-end=\"468\">ISO 7864<\/strong> defines the <em data-start=\"481\" data-end=\"558\">requirements and test methods for sterile hypodermic needles for single use<\/em>. This international standard ensures that medical needles used for injection or fluid withdrawal meet stringent safety, sharpness, and performance requirements.<\/p>\n<p data-start=\"873\" data-end=\"1157\">Op <a href=\"https:\/\/www.qualitester.com\/nl\/\" target=\"_blank\" rel=\"noopener\"><strong data-start=\"876\" data-end=\"906\">Cell Instruments Co.<\/strong><\/a>, we provide advanced solutions for testing the mechanical properties and functionality of medical needles and syringes in accordance with ISO 7864. Our instruments are widely used in medical device quality control, research, and compliance testing.<\/p>\n<hr data-start=\"1159\" data-end=\"1162\" \/>\n<h2 data-start=\"1164\" data-end=\"1211\">Understanding ISO 7864 and Its Importance<\/h2>\n<p data-start=\"1212\" data-end=\"1513\"><strong data-start=\"1212\" data-end=\"1224\">ISO 7864<\/strong> establishes uniform specifications to ensure the <em data-start=\"1274\" data-end=\"1321\">sterility, integrity, and mechanical strength<\/em> of hypodermic needles. These standards are critical for maintaining patient safety by minimizing insertion pain, preventing breakage, and ensuring proper bonding between the needle and hub.<\/p>\n<p data-start=\"1515\" data-end=\"1794\">The standard outlines detailed procedures for various <strong data-start=\"1569\" data-end=\"1598\">hypodermic needle testing<\/strong> parameters, such as <em data-start=\"1619\" data-end=\"1671\">needle sharpness, penetrability, bonding strength,<\/em> en <em data-start=\"1676\" data-end=\"1699\">dimensional accuracy.<\/em> Each of these tests helps evaluate how effectively a needle can perform during clinical use.<\/p>\n<hr data-start=\"1796\" data-end=\"1799\" \/>\n<h2 data-start=\"1801\" data-end=\"1835\">Key ISO 7864 Testing Methods<\/h2>\n<h3 data-start=\"1837\" data-end=\"1905\">Annex D &#8211; Hypodermic Needle Sharpness Testing and Penetrability Testing<\/h3>\n<p data-start=\"1906\" data-end=\"2148\">One of the most important performance factors in <strong data-start=\"1955\" data-end=\"1984\">hypodermic needle testing<\/strong> is its sharpness. <strong data-start=\"2003\" data-end=\"2042\">Hypodermic needle sharpness testing<\/strong> evaluates the <em data-start=\"2057\" data-end=\"2091\">penetration force and drag force<\/em> during needle insertion into a standardized substrate.<\/p>\n<p data-start=\"2150\" data-end=\"2376\">According to ISO 7864 Annex D, the <strong data-start=\"2185\" data-end=\"2217\">needle penetrability testing<\/strong> is performed by inserting the needle into a substrate (often latex or polyurethane film) at a constant speed of <strong data-start=\"2330\" data-end=\"2344\">100 mm\/min<\/strong>. Two parameters are measured:<\/p>\n<ul>\n<li data-start=\"2379\" data-end=\"2461\"><strong data-start=\"2379\" data-end=\"2405\">Peak penetration force<\/strong> \u2013 the maximum force required to pierce the substrate.<\/li>\n<li data-start=\"2464\" data-end=\"2559\"><strong data-start=\"2464\" data-end=\"2478\">Drag force<\/strong> \u2013 the average frictional resistance as the needle moves through the substrate.<\/li>\n<\/ul>\n<p data-start=\"2561\" data-end=\"2705\">These values are plotted in a force\u2013displacement curve, representing both the initial puncture and the frictional forces acting on the needle.<\/p>\n<p data-start=\"2707\" data-end=\"2943\">To perform this test accurately, <strong data-start=\"2740\" data-end=\"2794\">Cel-instrumenten\u2019 <a href=\"https:\/\/www.qualitester.com\/nl\/product\/iso-7864-needle-penetration-force-measurement-instrument\/\" target=\"_blank\" rel=\"noopener\">NPD-01 Naaldpenetratie- en sleepkrachttester<\/a><\/strong>\u00a0offers high-precision measurement of penetration and drag forces, enabling users to assess needle performance in compliance with ISO 7864 Annex D.<\/p>\n<h3 data-start=\"2950\" data-end=\"2985\">Annex E &#8211; Needle Bonding Strength Test<\/h3>\n<p data-start=\"2986\" data-end=\"3264\">Another critical part of <strong data-start=\"3011\" data-end=\"3031\">iso 7864 testing<\/strong> is evaluating the mechanical connection between the <em data-start=\"3084\" data-end=\"3097\">needle tube<\/em> en <em data-start=\"3102\" data-end=\"3114\">needle hub<\/em>. De <strong data-start=\"3120\" data-end=\"3152\">needle bonding strength test<\/strong> (Annex E) ensures that the needle can withstand the tensile forces encountered during use without detachment.<\/p>\n<p data-start=\"3266\" data-end=\"3537\">In this test, the needle is mounted vertically on a tensile testing machine. The load cell is set to zero, and the sample is pulled at a speed of <strong data-start=\"3412\" data-end=\"3425\">50 mm\/min<\/strong> until the hub separates or the needle breaks. The <strong data-start=\"3476\" data-end=\"3502\">maximum force recorded<\/strong> represents the bonding strength.<\/p>\n<p data-start=\"3539\" data-end=\"3831\">For reliable and repeatable results, <strong data-start=\"3576\" data-end=\"3635\">Cell Instruments\u2019 NBS-01 Needle Bonding Strength Tester<\/strong> provides accurate force measurement and high sampling precision. This instrument is designed specifically to evaluate bonding strength according to ISO 7864 and related standards like ISO 9626.<\/p>\n<hr data-start=\"3833\" data-end=\"3836\" \/>\n<h2 data-start=\"3838\" data-end=\"3897\">Syringe Functionality and Hypodermic Syringe Testing<\/h2>\n<p data-start=\"3898\" data-end=\"4189\">While ISO 7864 focuses on needles, it is often used alongside <a href=\"https:\/\/www.qualitester.com\/nl\/standards\/iso-7886-1\/\" target=\"_blank\" rel=\"noopener\"><strong data-start=\"3960\" data-end=\"3972\">ISO 7886-1<\/strong><\/a>, which covers <em data-start=\"3987\" data-end=\"4023\">hypodermic syringes for single use<\/em>. <strong data-start=\"4025\" data-end=\"4058\">Syringe functionality testing<\/strong> examines the ease of operation, plunger force, and leakage prevention\u2014ensuring the complete injection system functions reliably.<\/p>\n<p data-start=\"4191\" data-end=\"4379\">When combined with <strong data-start=\"4210\" data-end=\"4240\">hypodermic syringe testing<\/strong>, ISO 7864 testing provides a full evaluation of the <em data-start=\"4293\" data-end=\"4318\">needle-syringe assembly<\/em>, confirming that both components work seamlessly together.<\/p>\n<hr data-start=\"5026\" data-end=\"5029\" \/>\n<h2 data-start=\"5031\" data-end=\"5085\">Why Choose Cell Instruments for ISO 7864 Testing<\/h2>\n<p><img fetchpriority=\"high\" decoding=\"async\" class=\"size-medium wp-image-7570 alignleft\" src=\"https:\/\/www.qualitester.com\/wp-content\/uploads\/2025\/11\/ISO-7864-Sterile-Hypodermic-Needle-Penetration-and-Drag-Force-Needle-Bonding-Strength-300x300.webp\" alt=\"ISO 7864 Sterile Hypodermic Needle Penetration and Drag Force &amp; Needle Bonding Strength\" width=\"300\" height=\"300\" srcset=\"https:\/\/www.qualitester.com\/wp-content\/uploads\/2025\/11\/ISO-7864-Sterile-Hypodermic-Needle-Penetration-and-Drag-Force-Needle-Bonding-Strength-300x300.webp 300w, https:\/\/www.qualitester.com\/wp-content\/uploads\/2025\/11\/ISO-7864-Sterile-Hypodermic-Needle-Penetration-and-Drag-Force-Needle-Bonding-Strength-150x150.webp 150w, https:\/\/www.qualitester.com\/wp-content\/uploads\/2025\/11\/ISO-7864-Sterile-Hypodermic-Needle-Penetration-and-Drag-Force-Needle-Bonding-Strength-12x12.webp 12w, https:\/\/www.qualitester.com\/wp-content\/uploads\/2025\/11\/ISO-7864-Sterile-Hypodermic-Needle-Penetration-and-Drag-Force-Needle-Bonding-Strength-100x100.webp 100w, https:\/\/www.qualitester.com\/wp-content\/uploads\/2025\/11\/ISO-7864-Sterile-Hypodermic-Needle-Penetration-and-Drag-Force-Needle-Bonding-Strength.webp 600w\" sizes=\"(max-width: 300px) 100vw, 300px\" \/><\/p>\n<p data-start=\"5086\" data-end=\"5276\">With years of expertise in material and medical device testing, <strong data-start=\"5150\" data-end=\"5170\">Cel Instrumenten<\/strong> provides advanced, automated solutions that meet international standards such as ISO 7864 and ISO 9626.<\/p>\n<p data-start=\"5278\" data-end=\"5375\">Our <strong data-start=\"2740\" data-end=\"2794\">NPD-01 Naaldpenetratie- en sleepkrachttester<\/strong>\u00a0en <strong data-start=\"5323\" data-end=\"5364\">NBS-01 Needle Bonding Strength Tester<\/strong> deliver:<\/p>\n<ul>\n<li data-start=\"5378\" data-end=\"5427\">High accuracy load cells for force measurement.<\/li>\n<li data-start=\"5430\" data-end=\"5480\">Programmable test speeds and penetration depths.<\/li>\n<li data-start=\"5483\" data-end=\"5523\">Automated data recording and analysis.<\/li>\n<li data-start=\"5526\" data-end=\"5585\">User-friendly interfaces for efficient testing workflows.<\/li>\n<\/ul>\n<p data-start=\"5587\" data-end=\"5771\">These systems are ideal for <strong data-start=\"5615\" data-end=\"5692\">hypodermic needle manufacturers, syringe producers, medical laboratories,<\/strong> en <strong data-start=\"5697\" data-end=\"5725\">quality control agencies<\/strong> seeking ISO-compliant testing capabilities.<\/p>\n<hr data-start=\"5773\" data-end=\"5776\" \/>\n<p data-start=\"5795\" data-end=\"6108\"><a href=\"https:\/\/www.iso.org\/standard\/60481.html\" target=\"_blank\" rel=\"noopener nofollow\"><strong data-start=\"5795\" data-end=\"5807\">ISO 7864<\/strong><\/a> plays a vital role in ensuring the <em data-start=\"5843\" data-end=\"5881\">safety, performance, and reliability<\/em> of sterile hypodermic needles. Through standardized <strong data-start=\"5934\" data-end=\"6021\">needle sharpness tests, penetrability assessments, and bonding strength evaluations<\/strong>, manufacturers can deliver products that meet both regulatory and user expectations.<\/p>\n<p data-start=\"6110\" data-end=\"6383\">By using advanced testing instruments like <strong data-start=\"6153\" data-end=\"6163\">NPD-01<\/strong> en <strong data-start=\"6168\" data-end=\"6178\">NBS-01<\/strong> van <strong data-start=\"6184\" data-end=\"6204\">Cel Instrumenten<\/strong>, laboratories can achieve precise, repeatable results fully aligned with ISO 7864 requirements\u2014enhancing both <em data-start=\"6315\" data-end=\"6334\">quality assurance<\/em> en <em data-start=\"6339\" data-end=\"6355\">customer trust<\/em> in every needle produced.<\/p>","protected":false},"excerpt":{"rendered":"<p>ISO 7864 Sterile Hypodermic Needle Testing Penetration and Drag Force &amp; Needle Bonding Strength ISO 7864 defines the requirements and test methods for sterile hypodermic needles for single use. This international standard ensures that medical needles used for injection or fluid withdrawal meet stringent safety, sharpness, and performance requirements. At Cell Instruments Co., Ltd., we &hellip;<\/p>\n<p class=\"read-more\"> <a class=\"\" href=\"https:\/\/www.qualitester.com\/nl\/standards\/iso-7864\/\"> <span class=\"screen-reader-text\">ISO 7864<\/span> Lees meer \"<\/a><\/p>","protected":false},"author":4,"featured_media":0,"parent":4948,"menu_order":0,"comment_status":"closed","ping_status":"closed","template":"","meta":{"site-sidebar-layout":"default","site-content-layout":"default","ast-global-header-display":"","ast-main-header-display":"","ast-hfb-above-header-display":"","ast-hfb-below-header-display":"","ast-hfb-mobile-header-display":"","site-post-title":"disabled","ast-breadcrumbs-content":"","ast-featured-img":"disabled","footer-sml-layout":"","theme-transparent-header-meta":"default","adv-header-id-meta":"","stick-header-meta":"","header-above-stick-meta":"","header-main-stick-meta":"","header-below-stick-meta":"","footnotes":""},"class_list":["post-7567","page","type-page","status-publish","hentry"],"_links":{"self":[{"href":"https:\/\/www.qualitester.com\/nl\/wp-json\/wp\/v2\/pages\/7567","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/www.qualitester.com\/nl\/wp-json\/wp\/v2\/pages"}],"about":[{"href":"https:\/\/www.qualitester.com\/nl\/wp-json\/wp\/v2\/types\/page"}],"author":[{"embeddable":true,"href":"https:\/\/www.qualitester.com\/nl\/wp-json\/wp\/v2\/users\/4"}],"replies":[{"embeddable":true,"href":"https:\/\/www.qualitester.com\/nl\/wp-json\/wp\/v2\/comments?post=7567"}],"version-history":[{"count":0,"href":"https:\/\/www.qualitester.com\/nl\/wp-json\/wp\/v2\/pages\/7567\/revisions"}],"up":[{"embeddable":true,"href":"https:\/\/www.qualitester.com\/nl\/wp-json\/wp\/v2\/pages\/4948"}],"wp:attachment":[{"href":"https:\/\/www.qualitester.com\/nl\/wp-json\/wp\/v2\/media?parent=7567"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}