{"id":6648,"date":"2025-05-09T11:42:03","date_gmt":"2025-05-09T03:42:03","guid":{"rendered":"https:\/\/www.qualitester.com\/?page_id=6648"},"modified":"2025-11-13T09:38:29","modified_gmt":"2025-11-13T01:38:29","slug":"iso-7886-1","status":"publish","type":"page","link":"https:\/\/www.qualitester.com\/nb\/standards\/iso-7886-1\/","title":{"rendered":"ISO 7886-1"},"content":{"rendered":"<h1 style=\"text-align: center;\" data-start=\"370\" data-end=\"882\"><small>ISO 7886-1<\/small><\/h1>\n<h1 style=\"text-align: center;\" data-start=\"370\" data-end=\"882\"><small>Test med hypodermisk spr\u00f8yte<\/small><\/h1>\n<p class=\"\" data-start=\"370\" data-end=\"882\">Den <strong data-start=\"349\" data-end=\"363\">ISO 7886-1<\/strong> standarden definerer ytelses- og sikkerhetskravene for <em data-start=\"425\" data-end=\"446\">hypodermiske spr\u00f8yter<\/em> beregnet for engangsbruk, med vekt p\u00e5 aspekter som <em data-start=\"497\" data-end=\"514\">lekkasjemotstand<\/em>, <em data-start=\"516\" data-end=\"528\">d\u00f8dplass<\/em>, og <em data-start=\"534\" data-end=\"547\">glidekraft<\/em>. Produsenter og kvalitetssikringsteam m\u00e5 forst\u00e5 testprosedyrene som er beskrevet i ISO 7886-1 for \u00e5 sikre produktsamsvar og pasientsikkerhet.<\/p>\n<p class=\"\" data-start=\"1369\" data-end=\"1391\">ISO 7886-1 gjelder for:<\/p>\n<ul data-start=\"1393\" data-end=\"1559\">\n<li class=\"\" data-start=\"1393\" data-end=\"1438\">\n<p class=\"\" data-start=\"1395\" data-end=\"1438\">Sterile spr\u00f8yter laget av plastmaterialer.<\/p>\n<\/li>\n<li class=\"\" data-start=\"1439\" data-end=\"1523\">\n<p class=\"\" data-start=\"1441\" data-end=\"1523\">Spr\u00f8yter som brukes med kanyler for subkutan eller intramuskul\u00e6r injeksjon.<\/p>\n<\/li>\n<li class=\"\" data-start=\"1524\" data-end=\"1559\">\n<p class=\"\" data-start=\"1526\" data-end=\"1559\">Spr\u00f8yter beregnet for engangsbruk.<\/p>\n<\/li>\n<\/ul>\n<p class=\"\" data-start=\"1561\" data-end=\"1706\">Denne standarden gj\u00f8r <strong data-start=\"1580\" data-end=\"1593\">gjelder ikke<\/strong> til spr\u00f8yter for insulin (dekket av ISO 8537), tannlegebruk (dekket av ISO 7885) eller spr\u00f8yter til infusjonspumper.<\/p>\n<hr class=\"\" data-start=\"884\" data-end=\"887\" \/>\n<h2 class=\"\" data-start=\"912\" data-end=\"952\">ISO 7886-testing og dens betydning<\/h2>\n<p class=\"\" data-start=\"954\" data-end=\"1117\"><strong data-start=\"954\" data-end=\"968\">ISO 7886-1<\/strong> fokuserer p\u00e5 ytelseskrav og testmetoder for <em data-start=\"1026\" data-end=\"1047\">hypodermiske spr\u00f8yter<\/em>spesielt de som brukes med eller uten kanyler. Disse testene sikrer:<\/p>\n<h3 class=\"\" data-start=\"1861\" data-end=\"1895\">1. <strong data-start=\"1868\" data-end=\"1895\">Design og konstruksjon<\/strong><\/h3>\n<ul data-start=\"1896\" data-end=\"2054\">\n<li class=\"\" data-start=\"1896\" data-end=\"1951\">\n<p class=\"\" data-start=\"1898\" data-end=\"1951\">Gjennomsiktig fat for visuell inspeksjon av innholdet.<\/p>\n<\/li>\n<li class=\"\" data-start=\"1952\" data-end=\"1997\">\n<p class=\"\" data-start=\"1954\" data-end=\"1997\">Gradert skala trykt med permanent blekk.<\/p>\n<\/li>\n<li class=\"\" data-start=\"1998\" data-end=\"2054\">\n<p class=\"\" data-start=\"2000\" data-end=\"2054\">Sikker passform p\u00e5 stempelet for \u00e5 forhindre lekkasje eller at det l\u00f8sner.<\/p>\n<\/li>\n<\/ul>\n<h3 class=\"\" data-start=\"2056\" data-end=\"2082\">2. <strong data-start=\"2063\" data-end=\"2082\">Volumn\u00f8yaktighet<\/strong><\/h3>\n<ul data-start=\"2083\" data-end=\"2260\">\n<li class=\"\" data-start=\"2083\" data-end=\"2192\">\n<p class=\"\" data-start=\"2085\" data-end=\"2192\">Det faktiske leverte volumet m\u00e5 stemme overens med det nominelle volumet som er merket p\u00e5 fatet, innenfor spesifiserte toleranser.<\/p>\n<\/li>\n<li class=\"\" data-start=\"2193\" data-end=\"2260\">\n<p class=\"\" data-start=\"2195\" data-end=\"2260\">N\u00f8yaktigheten er testet ved 10%, 50% og 100% av den nominelle kapasiteten.<\/p>\n<\/li>\n<\/ul>\n<h3 class=\"\" data-start=\"2262\" data-end=\"2286\">3. <strong data-start=\"2269\" data-end=\"2286\">Lekkasjetester<\/strong><\/h3>\n<ul data-start=\"2287\" data-end=\"2457\">\n<li class=\"\" data-start=\"2287\" data-end=\"2384\">\n<p class=\"\" data-start=\"2289\" data-end=\"2384\"><strong data-start=\"2289\" data-end=\"2304\">Luftlekkasje<\/strong> og <strong data-start=\"2309\" data-end=\"2327\">v\u00e6skelekkasje<\/strong> testes under definerte trykk- og volumforhold.<\/p>\n<\/li>\n<li class=\"\" data-start=\"2385\" data-end=\"2457\">\n<p class=\"\" data-start=\"2387\" data-end=\"2457\">Det skal ikke forekomme synlig lekkasje fra fatet eller mellom komponentene.<\/p>\n<\/li>\n<\/ul>\n<h3 class=\"\" data-start=\"2459\" data-end=\"2480\">4. <strong data-start=\"2466\" data-end=\"2480\">Dead Space<\/strong><\/h3>\n<ul data-start=\"2481\" data-end=\"2614\">\n<li class=\"\" data-start=\"2481\" data-end=\"2614\">\n<p class=\"\" data-start=\"2483\" data-end=\"2614\">Volumet som er igjen i spr\u00f8yten etter at stempelet er trykket helt ned, m\u00e5 ikke overstige det angitte maksimumet (vanligvis \u2264 0,07 mL).<\/p>\n<\/li>\n<\/ul>\n<h3 class=\"\" data-start=\"2616\" data-end=\"2643\">5. <strong data-start=\"2623\" data-end=\"2643\">Stempelfunksjon<\/strong><\/h3>\n<ul data-start=\"2644\" data-end=\"2782\">\n<li class=\"\" data-start=\"2644\" data-end=\"2708\">\n<p class=\"\" data-start=\"2646\" data-end=\"2708\">Stempelet skal bevege seg jevnt uten \u00e5 sette seg fast eller skli.<\/p>\n<\/li>\n<li class=\"\" data-start=\"2709\" data-end=\"2782\">\n<p class=\"\" data-start=\"2711\" data-end=\"2782\">Det m\u00e5 ikke forekomme lekkasje eller separasjon under normalt driftstrykk.<\/p>\n<\/li>\n<\/ul>\n<h3 class=\"\" data-start=\"2784\" data-end=\"2827\">6. <strong data-start=\"2791\" data-end=\"2827\">Bruddkraft og separasjonskraft<\/strong><\/h3>\n<ul data-start=\"2828\" data-end=\"2945\">\n<li class=\"\" data-start=\"2828\" data-end=\"2945\">\n<p class=\"\" data-start=\"2830\" data-end=\"2945\">Den maksimale kraften som kreves for \u00e5 knekke stempelet eller skille komponentene fra hverandre, skal ikke ligge under sikkerhetsgrensen.<\/p>\n<\/li>\n<\/ul>\n<h3 class=\"\" data-start=\"2947\" data-end=\"2974\">7. <strong data-start=\"2954\" data-end=\"2974\">Biokompatibilitet<\/strong><\/h3>\n<ul data-start=\"2975\" data-end=\"3080\">\n<li class=\"\" data-start=\"2975\" data-end=\"3080\">\n<p class=\"\" data-start=\"2977\" data-end=\"3080\">Materialer som kommer i kontakt med legemidlet eller injeksjonsstedet, m\u00e5 best\u00e5 biokompatibilitetstester i henhold til ISO 10993-serien.<\/p>\n<\/li>\n<\/ul>\n<p class=\"\" data-start=\"1313\" data-end=\"1462\">Hver test bidrar til \u00e5 avgj\u00f8re om en spr\u00f8yte er trygg, p\u00e5litelig og har jevn ytelse.<\/p>\n<hr class=\"\" data-start=\"1766\" data-end=\"1769\" \/>\n<h2 class=\"\" data-start=\"1568\" data-end=\"1616\">Testmetoder for hypodermiske spr\u00f8yter i ISO 7886-1<\/h2>\n<h3 class=\"\" data-start=\"1618\" data-end=\"1673\">1. Lekkasjetest med hypodermisk spr\u00f8yte under vakuum (vedlegg B)<img fetchpriority=\"high\" decoding=\"async\" class=\"wp-image-6651 alignright\" src=\"https:\/\/www.qualitester.com\/wp-content\/uploads\/2025\/05\/ISO-7886-1-Annex-B.webp\" alt=\"ISO 7886-1 vedlegg B\" width=\"398\" height=\"327\" srcset=\"https:\/\/www.qualitester.com\/wp-content\/uploads\/2025\/05\/ISO-7886-1-Annex-B.webp 694w, https:\/\/www.qualitester.com\/wp-content\/uploads\/2025\/05\/ISO-7886-1-Annex-B-300x246.webp 300w, https:\/\/www.qualitester.com\/wp-content\/uploads\/2025\/05\/ISO-7886-1-Annex-B-15x12.webp 15w, https:\/\/www.qualitester.com\/wp-content\/uploads\/2025\/05\/ISO-7886-1-Annex-B-600x493.webp 600w\" sizes=\"(max-width: 398px) 100vw, 398px\" \/><\/h3>\n<p class=\"\" data-start=\"1675\" data-end=\"1866\">Lekkasje forbi stempelstoppen under undertrykk m\u00e5 testes for \u00e5 unng\u00e5 kontaminering eller luftinntrengning. Dette <strong data-start=\"1793\" data-end=\"1825\">lekkasjetest for hypodermiske spr\u00f8yter<\/strong> er avgj\u00f8rende for \u00e5 verifisere lufttettheten.<\/p>\n<ul data-start=\"1868\" data-end=\"2062\">\n<li class=\"\" data-start=\"1868\" data-end=\"1979\">\n<p class=\"\" data-start=\"1870\" data-end=\"1979\"><strong data-start=\"1870\" data-end=\"1881\">Metode:<\/strong> Fyll spr\u00f8yten med vann (minst 25% kapasitet), klem fast stempelet, og p\u00e5f\u00f8r -88 kPa vakuum.<\/p>\n<\/li>\n<li class=\"\" data-start=\"1980\" data-end=\"2062\">\n<p class=\"\" data-start=\"1982\" data-end=\"2062\"><strong data-start=\"1982\" data-end=\"1998\">Observasjon:<\/strong> Kontroller om det er luftlekkasjer eller om stempelproppen har l\u00f8snet.<\/p>\n<\/li>\n<\/ul>\n<p class=\"\" data-start=\"2064\" data-end=\"2229\"><strong data-start=\"2064\" data-end=\"2090\">Anbefalt utstyr:<\/strong><br data-start=\"2090\" data-end=\"2093\" \/>Cell Instruments\u2019 <strong data-start=\"2111\" data-end=\"2141\">SLT-02 <a href=\"https:\/\/www.qualitester.com\/nb\/produkt-2\/syringe-air-leakage-tester-iso-7886-1\/\">Testapparat for spr\u00f8yte lekkasjer<\/a><\/strong> gir presis vakuumregulering og automatiske m\u00e5linger, noe som sikrer at kravene i vedlegg B oppfylles.<\/p>\n<hr class=\"\" data-start=\"2231\" data-end=\"2234\" \/>\n<h3 class=\"\" data-start=\"2236\" data-end=\"2303\">2. Test av v\u00e6skelekkasje fra injeksjonsspr\u00f8yter under trykk (vedlegg D)<\/h3>\n<p class=\"\" data-start=\"2305\" data-end=\"2418\">Den <strong data-start=\"2309\" data-end=\"2348\">Test av v\u00e6skelekkasje fra injeksjonsspr\u00f8yte<\/strong> unders\u00f8ker spr\u00f8ytenes holdbarhet under aksial og lateral mekanisk belastning.<\/p>\n<ul data-start=\"2420\" data-end=\"2647\">\n<li class=\"\" data-start=\"2420\" data-end=\"2565\">\n<p class=\"\" data-start=\"2422\" data-end=\"2565\"><strong data-start=\"2422\" data-end=\"2433\">Metode:<\/strong> Etter at dysen er forseglet, p\u00e5f\u00f8res en sidekraft for \u00e5 avb\u00f8ye stempelet, etterfulgt av aksial kompresjon for \u00e5 simulere brukerens betjening.<\/p>\n<\/li>\n<li class=\"\" data-start=\"2566\" data-end=\"2647\">\n<p class=\"\" data-start=\"2568\" data-end=\"2647\"><strong data-start=\"2568\" data-end=\"2584\">Observasjon:<\/strong> Kontroller for v\u00e6skelekkasje ved trykk p\u00e5 200 kPa og 300 kPa.<\/p>\n<\/li>\n<\/ul>\n<p class=\"\" data-start=\"2649\" data-end=\"2835\"><strong data-start=\"2649\" data-end=\"2675\">Anbefalt utstyr:<\/strong><br data-start=\"2675\" data-end=\"2678\" \/>Bruk <strong data-start=\"2682\" data-end=\"2728\">SPPT-01 <a href=\"https:\/\/www.qualitester.com\/nb\/produkt-2\/syringe-liquid-leakage-tester\/\">Tetthetstester for spr\u00f8yter med overtrykk<\/a><\/strong>\u00a0fra Cell Instruments for \u00e5 p\u00e5f\u00f8re kontrollerte mekaniske krefter og overv\u00e5ke om det oppst\u00e5r brudd i tetningens integritet.<\/p>\n<hr class=\"\" data-start=\"2837\" data-end=\"2840\" \/>\n<h3 data-start=\"2842\" data-end=\"2907\">3. Test av spr\u00f8ytenes d\u00f8drom: Minimering av medisinsvinn (vedlegg C)<\/h3>\n<p class=\"\" data-start=\"2909\" data-end=\"3067\">D\u00f8drom betegner v\u00e6skevolumet som blir igjen i spr\u00f8yten etter fullstendig utt\u00f8mming. Et stort d\u00f8drom kan f\u00f8re til sl\u00f8sing med legemiddel og un\u00f8yaktig dosering.<\/p>\n<ul data-start=\"3069\" data-end=\"3238\">\n<li class=\"\" data-start=\"3069\" data-end=\"3162\">\n<p class=\"\" data-start=\"3071\" data-end=\"3162\"><strong data-start=\"3071\" data-end=\"3082\">Metode:<\/strong> Vei den tomme spr\u00f8yten, og vei den deretter igjen etter at den er fylt med vann og t\u00f8mt for vann.<\/p>\n<\/li>\n<li class=\"\" data-start=\"3163\" data-end=\"3238\">\n<p class=\"\" data-start=\"3165\" data-end=\"3238\"><strong data-start=\"3165\" data-end=\"3181\">Beregning:<\/strong> Vektforskjellen tyder p\u00e5 <strong data-start=\"3217\" data-end=\"3231\">d\u00f8dplass<\/strong> i ml.<\/p>\n<\/li>\n<\/ul>\n<hr class=\"\" data-start=\"3437\" data-end=\"3440\" \/>\n<h3 data-start=\"3442\" data-end=\"3490\">4. Glidekraftpr\u00f8ve for spr\u00f8ytekolv (vedlegg E)<img decoding=\"async\" class=\"wp-image-6652 alignright\" src=\"https:\/\/www.qualitester.com\/wp-content\/uploads\/2025\/05\/ISO-7886-1-Annex-E.webp\" alt=\"ISO 7886-1, vedlegg E\" width=\"272\" height=\"358\" srcset=\"https:\/\/www.qualitester.com\/wp-content\/uploads\/2025\/05\/ISO-7886-1-Annex-E.webp 546w, https:\/\/www.qualitester.com\/wp-content\/uploads\/2025\/05\/ISO-7886-1-Annex-E-228x300.webp 228w, https:\/\/www.qualitester.com\/wp-content\/uploads\/2025\/05\/ISO-7886-1-Annex-E-9x12.webp 9w\" sizes=\"(max-width: 272px) 100vw, 272px\" \/><\/h3>\n<p class=\"\" data-start=\"3492\" data-end=\"3683\">Jevn stempelbevegelse er avgj\u00f8rende for brukerkomfort og presis doseringskontroll. Den <strong data-start=\"3576\" data-end=\"3596\">glidekraftpr\u00f8ve<\/strong> m\u00e5ler kraften som kreves for \u00e5 drive stempelet jevnt gjennom hele slaget.<\/p>\n<ul data-start=\"3685\" data-end=\"3873\">\n<li class=\"\" data-start=\"3685\" data-end=\"3798\">\n<p class=\"\" data-start=\"3687\" data-end=\"3798\"><strong data-start=\"3687\" data-end=\"3698\">Metode:<\/strong> En mekanisk testmaskin trykker stempelet ned med en fast hastighet og registrerer motstanden i sanntid.<\/p>\n<\/li>\n<li class=\"\" data-start=\"3799\" data-end=\"3873\">\n<p class=\"\" data-start=\"3801\" data-end=\"3873\"><strong data-start=\"3801\" data-end=\"3813\">Form\u00e5l:<\/strong> Kontrollerer at spr\u00f8ytene er brukbare og at kvaliteten er jevn p\u00e5 tvers av partiene.<\/p>\n<\/li>\n<\/ul>\n<p class=\"\" data-start=\"3875\" data-end=\"4084\"><strong data-start=\"3875\" data-end=\"3901\">Anbefalt utstyr:<\/strong><br data-start=\"3901\" data-end=\"3904\" \/><strong data-start=\"3904\" data-end=\"3958\">Cell Instruments MST-01 <a href=\"https:\/\/www.qualitester.com\/nb\/produkt-2\/sale-syringe-plunger-glide-force-tester\/\">Medisinsk spr\u00f8ytetester<\/a><\/strong>\u00a0er ideelt for <strong data-start=\"3972\" data-end=\"3992\">glidekraftpr\u00f8ve<\/strong> applikasjoner. Den registrerer kraftprofiler og oppfyller n\u00f8yaktighetskravene i ISO 7886-1.<\/p>\n<hr class=\"\" data-start=\"4086\" data-end=\"4089\" \/>\n<h2 class=\"\" data-start=\"4091\" data-end=\"4143\">Hvorfor velge Cell Instruments for testing i henhold til ISO 7886?<\/h2>\n<p class=\"\" data-start=\"4145\" data-end=\"4312\">Cell Instruments tilbyr et komplett utvalg av l\u00f8sninger for testing av spr\u00f8yter, som hjelper produsenter med \u00e5 oppfylle og overg\u00e5 <strong data-start=\"4253\" data-end=\"4267\">ISO 7886-1<\/strong> krav. Utstyret v\u00e5rt er utviklet for:<\/p>\n<ul data-start=\"4313\" data-end=\"4476\">\n<li class=\"\" data-start=\"4313\" data-end=\"4344\">\n<p class=\"\" data-start=\"4315\" data-end=\"4344\">H\u00f8y presisjon og automatisering<\/p>\n<\/li>\n<li class=\"\" data-start=\"4345\" data-end=\"4386\">\n<p class=\"\" data-start=\"4347\" data-end=\"4386\">Overholdelse av internasjonale standarder<\/p>\n<\/li>\n<li class=\"\" data-start=\"4387\" data-end=\"4422\">\n<p class=\"\" data-start=\"4389\" data-end=\"4422\">Enkel betjening og dataregistrering<\/p>\n<\/li>\n<li class=\"\" data-start=\"4423\" data-end=\"4476\">\n<p class=\"\" data-start=\"4425\" data-end=\"4476\">Fleksibel tilpasning for \u00e5 im\u00f8tekomme spesifikke testoppsett<\/p>\n<\/li>\n<\/ul>\n<p class=\"\" data-start=\"4478\" data-end=\"4682\">Uansett om du skal validere produktdesign eller administrere testing i forbindelse med frigivelse av produksjonspartier, <strong data-start=\"4554\" data-end=\"4564\">SLT-01<\/strong>, <strong data-start=\"4566\" data-end=\"4577\">SPPT-01<\/strong>, og <strong data-start=\"4583\" data-end=\"4593\">MST-01<\/strong> levere p\u00e5litelige og repeterbare resultater som er tilpasset behovene i medisinsk utstyrsbransjen.<\/p>\n<hr class=\"\" data-start=\"4684\" data-end=\"4687\" \/>\n<p class=\"\" data-start=\"4704\" data-end=\"5183\">\u00c5 oppn\u00e5 samsvar med <a href=\"https:\/\/www.iso.org\/standard\/64790.html\" target=\"_blank\" rel=\"noopener nofollow\"><strong data-start=\"4730\" data-end=\"4744\">ISO 7886-1<\/strong><\/a> er ikke bare et regulatorisk hinder \u2013 det er en forpliktelse til kvalitet og pasientsikkerhet. Ved \u00e5 forst\u00e5 og gjennomf\u00f8re viktige tester som <strong data-start=\"4885\" data-end=\"4912\">test med injeksjonsspr\u00f8yte<\/strong>, <strong data-start=\"4914\" data-end=\"4927\">lekkasjetest<\/strong>, <strong data-start=\"4929\" data-end=\"4956\">Test av spr\u00f8ytenes d\u00f8drom<\/strong>, og <strong data-start=\"4962\" data-end=\"4982\">glidekraftpr\u00f8ve<\/strong>, kan produsentene trygt lansere produktene sine p\u00e5 markedet. Utstyr laboratoriet ditt med Cell Instruments\u2019 avanserte testutstyr for \u00e5 effektivisere testingen og sikre n\u00f8yaktighet i alle ledd av spr\u00f8yteproduksjonen.<\/p>\n<p class=\"\" data-start=\"5185\" data-end=\"5296\"><strong data-start=\"5185\" data-end=\"5296\">For teknisk r\u00e5dgivning, valg av instrumenter eller skreddersydde l\u00f8sninger, ta kontakt med Cell Instruments i dag.<\/strong><\/p>","protected":false},"excerpt":{"rendered":"<p>ISO 7886-1-test av injeksjonsspr\u00f8yter: ISO 7886-1-standarden definerer ytelses- og sikkerhetskravene for injeksjonsspr\u00f8yter beregnet p\u00e5 engangsbruk, med vekt p\u00e5 aspekter som lekkasjebestandighet, d\u00f8drom og glidekraft. Produsenter og kvalitetssikringsteam m\u00e5 forst\u00e5 testprosedyrene som er beskrevet i ISO 7886-1 for \u00e5 sikre at produktene oppfyller kravene og pasientsikkerheten. ISO 7886-1 \u2026<\/p>\n<p class=\"read-more\"> <a class=\"\" href=\"https:\/\/www.qualitester.com\/nb\/standards\/iso-7886-1\/\"> <span class=\"screen-reader-text\">ISO 7886-1<\/span> Les mer \"<\/a><\/p>","protected":false},"author":4,"featured_media":0,"parent":4948,"menu_order":0,"comment_status":"closed","ping_status":"closed","template":"","meta":{"site-sidebar-layout":"default","site-content-layout":"default","ast-global-header-display":"","ast-main-header-display":"","ast-hfb-above-header-display":"","ast-hfb-below-header-display":"","ast-hfb-mobile-header-display":"","site-post-title":"disabled","ast-breadcrumbs-content":"","ast-featured-img":"disabled","footer-sml-layout":"","theme-transparent-header-meta":"default","adv-header-id-meta":"","stick-header-meta":"","header-above-stick-meta":"","header-main-stick-meta":"","header-below-stick-meta":"","footnotes":""},"class_list":["post-6648","page","type-page","status-publish","hentry"],"_links":{"self":[{"href":"https:\/\/www.qualitester.com\/nb\/wp-json\/wp\/v2\/pages\/6648","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/www.qualitester.com\/nb\/wp-json\/wp\/v2\/pages"}],"about":[{"href":"https:\/\/www.qualitester.com\/nb\/wp-json\/wp\/v2\/types\/page"}],"author":[{"embeddable":true,"href":"https:\/\/www.qualitester.com\/nb\/wp-json\/wp\/v2\/users\/4"}],"replies":[{"embeddable":true,"href":"https:\/\/www.qualitester.com\/nb\/wp-json\/wp\/v2\/comments?post=6648"}],"version-history":[{"count":0,"href":"https:\/\/www.qualitester.com\/nb\/wp-json\/wp\/v2\/pages\/6648\/revisions"}],"up":[{"embeddable":true,"href":"https:\/\/www.qualitester.com\/nb\/wp-json\/wp\/v2\/pages\/4948"}],"wp:attachment":[{"href":"https:\/\/www.qualitester.com\/nb\/wp-json\/wp\/v2\/media?parent=6648"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}