{"id":6648,"date":"2025-05-09T11:42:03","date_gmt":"2025-05-09T03:42:03","guid":{"rendered":"https:\/\/www.qualitester.com\/?page_id=6648"},"modified":"2025-11-13T09:38:29","modified_gmt":"2025-11-13T01:38:29","slug":"iso-7886-1","status":"publish","type":"page","link":"https:\/\/www.qualitester.com\/da\/standards\/iso-7886-1\/","title":{"rendered":"ISO 7886-1"},"content":{"rendered":"<h1 style=\"text-align: center;\" data-start=\"370\" data-end=\"882\"><small>ISO 7886-1<\/small><\/h1>\n<h1 style=\"text-align: center;\" data-start=\"370\" data-end=\"882\"><small>Test med injektionsspr\u00f8jte<\/small><\/h1>\n<p class=\"\" data-start=\"370\" data-end=\"882\">Den <strong data-start=\"349\" data-end=\"363\">ISO 7886-1<\/strong> Standarden fastl\u00e6gger ydeevne- og sikkerhedskravene for <em data-start=\"425\" data-end=\"446\">injektionsspr\u00f8jter<\/em> beregnet til engangsbrug, med fokus p\u00e5 aspekter som <em data-start=\"497\" data-end=\"514\">t\u00e6thed<\/em>, <em data-start=\"516\" data-end=\"528\">d\u00f8dplads<\/em>, og <em data-start=\"534\" data-end=\"547\">glidekraft<\/em>. Producenter og kvalitetssikringsteams skal have kendskab til de testprocedurer, der er beskrevet i ISO 7886-1, for at sikre, at produkterne overholder kravene, og at patientsikkerheden er i orden.<\/p>\n<p class=\"\" data-start=\"1369\" data-end=\"1391\">ISO 7886-1 g\u00e6lder for:<\/p>\n<ul data-start=\"1393\" data-end=\"1559\">\n<li class=\"\" data-start=\"1393\" data-end=\"1438\">\n<p class=\"\" data-start=\"1395\" data-end=\"1438\">Sterile spr\u00f8jter fremstillet af plastmaterialer.<\/p>\n<\/li>\n<li class=\"\" data-start=\"1439\" data-end=\"1523\">\n<p class=\"\" data-start=\"1441\" data-end=\"1523\">Spr\u00f8jter, der anvendes sammen med injektionsn\u00e5le til subkutan eller intramuskul\u00e6r injektion.<\/p>\n<\/li>\n<li class=\"\" data-start=\"1524\" data-end=\"1559\">\n<p class=\"\" data-start=\"1526\" data-end=\"1559\">Spr\u00f8jter beregnet til engangsbrug.<\/p>\n<\/li>\n<\/ul>\n<p class=\"\" data-start=\"1561\" data-end=\"1706\">Denne standard g\u00f8r f\u00f8lgende: <strong data-start=\"1580\" data-end=\"1593\">g\u00e6lder ikke<\/strong> til insulinspr\u00f8jter (omfattet af ISO 8537), spr\u00f8jter til tandl\u00e6gebrug (omfattet af ISO 7885) eller spr\u00f8jter til infusionspumper.<\/p>\n<hr class=\"\" data-start=\"884\" data-end=\"887\" \/>\n<h2 class=\"\" data-start=\"912\" data-end=\"952\">ISO 7886-pr\u00f8vning og dens betydning<\/h2>\n<p class=\"\" data-start=\"954\" data-end=\"1117\"><strong data-start=\"954\" data-end=\"968\">ISO 7886-1<\/strong> fokuserer p\u00e5 ydeevnekrav og pr\u00f8vningsmetoder for <em data-start=\"1026\" data-end=\"1047\">injektionsspr\u00f8jter<\/em>, is\u00e6r dem, der anvendes med eller uden n\u00e5le. Disse test sikrer:<\/p>\n<h3 class=\"\" data-start=\"1861\" data-end=\"1895\">1. <strong data-start=\"1868\" data-end=\"1895\">Design og konstruktion<\/strong><\/h3>\n<ul data-start=\"1896\" data-end=\"2054\">\n<li class=\"\" data-start=\"1896\" data-end=\"1951\">\n<p class=\"\" data-start=\"1898\" data-end=\"1951\">Gennemsigtig beholder, der g\u00f8r det muligt at se indholdet.<\/p>\n<\/li>\n<li class=\"\" data-start=\"1952\" data-end=\"1997\">\n<p class=\"\" data-start=\"1954\" data-end=\"1997\">Graderet skala trykt med permanent bl\u00e6k.<\/p>\n<\/li>\n<li class=\"\" data-start=\"1998\" data-end=\"2054\">\n<p class=\"\" data-start=\"2000\" data-end=\"2054\">S\u00f8rg for, at stemplet sidder fast, s\u00e5 der ikke opst\u00e5r l\u00e6kage eller l\u00f8sner sig.<\/p>\n<\/li>\n<\/ul>\n<h3 class=\"\" data-start=\"2056\" data-end=\"2082\">2. <strong data-start=\"2063\" data-end=\"2082\">M\u00e6ngden\u00f8jagtighed<\/strong><\/h3>\n<ul data-start=\"2083\" data-end=\"2260\">\n<li class=\"\" data-start=\"2083\" data-end=\"2192\">\n<p class=\"\" data-start=\"2085\" data-end=\"2192\">Den faktisk leverede m\u00e6ngde skal svare til den nominelle m\u00e6ngde, der er angivet p\u00e5 t\u00f8nden, inden for de angivne tolerancer.<\/p>\n<\/li>\n<li class=\"\" data-start=\"2193\" data-end=\"2260\">\n<p class=\"\" data-start=\"2195\" data-end=\"2260\">N\u00f8jagtigheden testes ved 10%, 50% og 100% af den nominelle kapacitet.<\/p>\n<\/li>\n<\/ul>\n<h3 class=\"\" data-start=\"2262\" data-end=\"2286\">3. <strong data-start=\"2269\" data-end=\"2286\">L\u00e6kagetests<\/strong><\/h3>\n<ul data-start=\"2287\" data-end=\"2457\">\n<li class=\"\" data-start=\"2287\" data-end=\"2384\">\n<p class=\"\" data-start=\"2289\" data-end=\"2384\"><strong data-start=\"2289\" data-end=\"2304\">Luftl\u00e6kage<\/strong> og <strong data-start=\"2309\" data-end=\"2327\">v\u00e6skel\u00e6kage<\/strong> testes under fastlagte tryk- og volumenbetingelser.<\/p>\n<\/li>\n<li class=\"\" data-start=\"2385\" data-end=\"2457\">\n<p class=\"\" data-start=\"2387\" data-end=\"2457\">Der m\u00e5 ikke forekomme synlige l\u00e6kager fra cylinderen eller mellem komponenterne.<\/p>\n<\/li>\n<\/ul>\n<h3 class=\"\" data-start=\"2459\" data-end=\"2480\">4. <strong data-start=\"2466\" data-end=\"2480\">Dead Space<\/strong><\/h3>\n<ul data-start=\"2481\" data-end=\"2614\">\n<li class=\"\" data-start=\"2481\" data-end=\"2614\">\n<p class=\"\" data-start=\"2483\" data-end=\"2614\">Det volumen, der er tilbage i spr\u00f8jten, efter at stemplet er trykket helt ned, m\u00e5 ikke overstige det angivne maksimum (normalt \u2264 0,07 ml).<\/p>\n<\/li>\n<\/ul>\n<h3 class=\"\" data-start=\"2616\" data-end=\"2643\">5. <strong data-start=\"2623\" data-end=\"2643\">Stempelets funktion<\/strong><\/h3>\n<ul data-start=\"2644\" data-end=\"2782\">\n<li class=\"\" data-start=\"2644\" data-end=\"2708\">\n<p class=\"\" data-start=\"2646\" data-end=\"2708\">Stemplet skal bev\u00e6ge sig j\u00e6vnt uden at s\u00e6tte sig fast eller glide.<\/p>\n<\/li>\n<li class=\"\" data-start=\"2709\" data-end=\"2782\">\n<p class=\"\" data-start=\"2711\" data-end=\"2782\">Der m\u00e5 ikke forekomme l\u00e6kage eller adskillelse ved normalt driftstryk.<\/p>\n<\/li>\n<\/ul>\n<h3 class=\"\" data-start=\"2784\" data-end=\"2827\">6. <strong data-start=\"2791\" data-end=\"2827\">Brudkraft og adskillelseskraft<\/strong><\/h3>\n<ul data-start=\"2828\" data-end=\"2945\">\n<li class=\"\" data-start=\"2828\" data-end=\"2945\">\n<p class=\"\" data-start=\"2830\" data-end=\"2945\">Den maksimale kraft, der kr\u00e6ves for at br\u00e6kke stemplet eller adskille komponenterne, b\u00f8r ikke ligge under sikkerhedst\u00e6rsklerne.<\/p>\n<\/li>\n<\/ul>\n<h3 class=\"\" data-start=\"2947\" data-end=\"2974\">7. <strong data-start=\"2954\" data-end=\"2974\">Biokompatibilitet<\/strong><\/h3>\n<ul data-start=\"2975\" data-end=\"3080\">\n<li class=\"\" data-start=\"2975\" data-end=\"3080\">\n<p class=\"\" data-start=\"2977\" data-end=\"3080\">Materialer, der kommer i kontakt med l\u00e6gemidlet eller injektionsstedet, skal best\u00e5 biokompatibilitetstestene i ISO 10993-serien.<\/p>\n<\/li>\n<\/ul>\n<p class=\"\" data-start=\"1313\" data-end=\"1462\">Hver test bidrager til at fastsl\u00e5, om en spr\u00f8jte er sikker, p\u00e5lidelig og yder ensartet.<\/p>\n<hr class=\"\" data-start=\"1766\" data-end=\"1769\" \/>\n<h2 class=\"\" data-start=\"1568\" data-end=\"1616\">Pr\u00f8vningsmetoder for injektionsspr\u00f8jter i ISO 7886-1<\/h2>\n<h3 class=\"\" data-start=\"1618\" data-end=\"1673\">1. L\u00e6kagetest af injektionsspr\u00f8jter under vakuum (bilag B)<img fetchpriority=\"high\" decoding=\"async\" class=\"wp-image-6651 alignright\" src=\"https:\/\/www.qualitester.com\/wp-content\/uploads\/2025\/05\/ISO-7886-1-Annex-B.webp\" alt=\"ISO 7886-1, bilag B\" width=\"398\" height=\"327\" srcset=\"https:\/\/www.qualitester.com\/wp-content\/uploads\/2025\/05\/ISO-7886-1-Annex-B.webp 694w, https:\/\/www.qualitester.com\/wp-content\/uploads\/2025\/05\/ISO-7886-1-Annex-B-300x246.webp 300w, https:\/\/www.qualitester.com\/wp-content\/uploads\/2025\/05\/ISO-7886-1-Annex-B-15x12.webp 15w, https:\/\/www.qualitester.com\/wp-content\/uploads\/2025\/05\/ISO-7886-1-Annex-B-600x493.webp 600w\" sizes=\"(max-width: 398px) 100vw, 398px\" \/><\/h3>\n<p class=\"\" data-start=\"1675\" data-end=\"1866\">Der skal foretages en test af l\u00e6kage forbi stempelproppen under undertryk for at undg\u00e5 forurening eller indtr\u00e6ngning af luft. Dette <strong data-start=\"1793\" data-end=\"1825\">L\u00e6kagetest af injektionsspr\u00f8jter<\/strong> er afg\u00f8rende for at kontrollere luftt\u00e6theden.<\/p>\n<ul data-start=\"1868\" data-end=\"2062\">\n<li class=\"\" data-start=\"1868\" data-end=\"1979\">\n<p class=\"\" data-start=\"1870\" data-end=\"1979\"><strong data-start=\"1870\" data-end=\"1881\">Metode:<\/strong> Fyld spr\u00f8jten med vand (mindst 25% kapacitet), fastklem stemplet, og p\u00e5f\u00f8r et vakuum p\u00e5 -88 kPa.<\/p>\n<\/li>\n<li class=\"\" data-start=\"1980\" data-end=\"2062\">\n<p class=\"\" data-start=\"1982\" data-end=\"2062\"><strong data-start=\"1982\" data-end=\"1998\">Bem\u00e6rkning:<\/strong> Kontroller, om der er luftl\u00e6kager, eller om stempelproppen har l\u00f8snet sig.<\/p>\n<\/li>\n<\/ul>\n<p class=\"\" data-start=\"2064\" data-end=\"2229\"><strong data-start=\"2064\" data-end=\"2090\">Anbefalet udstyr:<\/strong><br data-start=\"2090\" data-end=\"2093\" \/>Cell Instruments\u2019 <strong data-start=\"2111\" data-end=\"2141\">SLT-02 <a href=\"https:\/\/www.qualitester.com\/da\/vare\/syringe-air-leakage-tester-iso-7886-1\/\">Spr\u00f8jtel\u00e6kagetester<\/a><\/strong> giver mulighed for pr\u00e6cis vakuumregulering og automatiske m\u00e5linger, hvilket sikrer overholdelse af bilag B.<\/p>\n<hr class=\"\" data-start=\"2231\" data-end=\"2234\" \/>\n<h3 class=\"\" data-start=\"2236\" data-end=\"2303\">2. Test af v\u00e6skel\u00e6kage fra injektionsspr\u00f8jter under tryk (bilag D)<\/h3>\n<p class=\"\" data-start=\"2305\" data-end=\"2418\">Den <strong data-start=\"2309\" data-end=\"2348\">Test af v\u00e6skel\u00e6kage i injektionsspr\u00f8jter<\/strong> unders\u00f8ger spr\u00f8jtens holdbarhed under aksial og lateral mekanisk belastning.<\/p>\n<ul data-start=\"2420\" data-end=\"2647\">\n<li class=\"\" data-start=\"2420\" data-end=\"2565\">\n<p class=\"\" data-start=\"2422\" data-end=\"2565\"><strong data-start=\"2422\" data-end=\"2433\">Metode:<\/strong> Efter at dysen er blevet forseglet, p\u00e5f\u00f8res der en sidekraft for at afb\u00f8je stemplet, efterfulgt af en aksial kompression for at simulere brugerens betjening.<\/p>\n<\/li>\n<li class=\"\" data-start=\"2566\" data-end=\"2647\">\n<p class=\"\" data-start=\"2568\" data-end=\"2647\"><strong data-start=\"2568\" data-end=\"2584\">Bem\u00e6rkning:<\/strong> Kontroller for v\u00e6skel\u00e6kage ved tryk p\u00e5 200 kPa og 300 kPa.<\/p>\n<\/li>\n<\/ul>\n<p class=\"\" data-start=\"2649\" data-end=\"2835\"><strong data-start=\"2649\" data-end=\"2675\">Anbefalet udstyr:<\/strong><br data-start=\"2675\" data-end=\"2678\" \/>Brug <strong data-start=\"2682\" data-end=\"2728\">SPPT-01 <a href=\"https:\/\/www.qualitester.com\/da\/vare\/syringe-liquid-leakage-tester\/\">T\u00e6thedstester til spr\u00f8jter med overtryk<\/a><\/strong>\u00a0fra Cell Instruments til at p\u00e5f\u00f8re kontrollerede mekaniske kr\u00e6fter og overv\u00e5ge, om t\u00e6tningen er intakt.<\/p>\n<hr class=\"\" data-start=\"2837\" data-end=\"2840\" \/>\n<h3 data-start=\"2842\" data-end=\"2907\">3. Test af spr\u00f8jtens d\u00f8drum: Minimering af medicinspild (bilag C)<\/h3>\n<p class=\"\" data-start=\"2909\" data-end=\"3067\">D\u00f8drummet er den v\u00e6skem\u00e6ngde, der forbliver i spr\u00f8jten efter fuldst\u00e6ndig udt\u00f8mning. Et stort d\u00f8drum kan f\u00f8re til spild af l\u00e6gemiddel og un\u00f8jagtig dosering.<\/p>\n<ul data-start=\"3069\" data-end=\"3238\">\n<li class=\"\" data-start=\"3069\" data-end=\"3162\">\n<p class=\"\" data-start=\"3071\" data-end=\"3162\"><strong data-start=\"3071\" data-end=\"3082\">Metode:<\/strong> Vej den tomme spr\u00f8jte, og vej den derefter igen, efter at den er blevet fyldt med vand og t\u00f8mt for vand.<\/p>\n<\/li>\n<li class=\"\" data-start=\"3163\" data-end=\"3238\">\n<p class=\"\" data-start=\"3165\" data-end=\"3238\"><strong data-start=\"3165\" data-end=\"3181\">Beregning:<\/strong> V\u00e6gtforskellen viser, at <strong data-start=\"3217\" data-end=\"3231\">d\u00f8dplads<\/strong> i ml.<\/p>\n<\/li>\n<\/ul>\n<hr class=\"\" data-start=\"3437\" data-end=\"3440\" \/>\n<h3 data-start=\"3442\" data-end=\"3490\">4. Glidekraftpr\u00f8vning af spr\u00f8jtestemplet (bilag E)<img decoding=\"async\" class=\"wp-image-6652 alignright\" src=\"https:\/\/www.qualitester.com\/wp-content\/uploads\/2025\/05\/ISO-7886-1-Annex-E.webp\" alt=\"ISO 7886-1, bilag E\" width=\"272\" height=\"358\" srcset=\"https:\/\/www.qualitester.com\/wp-content\/uploads\/2025\/05\/ISO-7886-1-Annex-E.webp 546w, https:\/\/www.qualitester.com\/wp-content\/uploads\/2025\/05\/ISO-7886-1-Annex-E-228x300.webp 228w, https:\/\/www.qualitester.com\/wp-content\/uploads\/2025\/05\/ISO-7886-1-Annex-E-9x12.webp 9w\" sizes=\"(max-width: 272px) 100vw, 272px\" \/><\/h3>\n<p class=\"\" data-start=\"3492\" data-end=\"3683\">En j\u00e6vn stempelbev\u00e6gelse er afg\u00f8rende for brugerkomforten og en pr\u00e6cis doseringsstyring. Den <strong data-start=\"3576\" data-end=\"3596\">glidekraftstest<\/strong> m\u00e5ler den kraft, der kr\u00e6ves for at bev\u00e6ge stemplet j\u00e6vnt gennem hele slaget.<\/p>\n<ul data-start=\"3685\" data-end=\"3873\">\n<li class=\"\" data-start=\"3685\" data-end=\"3798\">\n<p class=\"\" data-start=\"3687\" data-end=\"3798\"><strong data-start=\"3687\" data-end=\"3698\">Metode:<\/strong> En mekanisk testmaskine trykker stemplet ned med en fast hastighed og registrerer modstanden i realtid.<\/p>\n<\/li>\n<li class=\"\" data-start=\"3799\" data-end=\"3873\">\n<p class=\"\" data-start=\"3801\" data-end=\"3873\"><strong data-start=\"3801\" data-end=\"3813\">Form\u00e5l:<\/strong> Kontrollerer, at spr\u00f8jterne er brugbare, og at der er ensartethed p\u00e5 tv\u00e6rs af partier.<\/p>\n<\/li>\n<\/ul>\n<p class=\"\" data-start=\"3875\" data-end=\"4084\"><strong data-start=\"3875\" data-end=\"3901\">Anbefalet udstyr:<\/strong><br data-start=\"3901\" data-end=\"3904\" \/><strong data-start=\"3904\" data-end=\"3958\">Cell Instruments MST-01 <a href=\"https:\/\/www.qualitester.com\/da\/vare\/sale-syringe-plunger-glide-force-tester\/\">Testapparat til medicinske spr\u00f8jter<\/a><\/strong>\u00a0er ideel til <strong data-start=\"3972\" data-end=\"3992\">glidekraftstest<\/strong> anvendelser. Den registrerer kraftprofiler og opfylder n\u00f8jagtighedskravene i ISO 7886-1.<\/p>\n<hr class=\"\" data-start=\"4086\" data-end=\"4089\" \/>\n<h2 class=\"\" data-start=\"4091\" data-end=\"4143\">Hvorfor v\u00e6lge Cell Instruments til ISO 7886-test?<\/h2>\n<p class=\"\" data-start=\"4145\" data-end=\"4312\">Cell Instruments tilbyder et komplet udvalg af l\u00f8sninger til test af spr\u00f8jter, der hj\u00e6lper producenter med at opfylde og overg\u00e5 <strong data-start=\"4253\" data-end=\"4267\">ISO 7886-1<\/strong> krav. Vores udstyr er udviklet til:<\/p>\n<ul data-start=\"4313\" data-end=\"4476\">\n<li class=\"\" data-start=\"4313\" data-end=\"4344\">\n<p class=\"\" data-start=\"4315\" data-end=\"4344\">H\u00f8j pr\u00e6cision og automatisering<\/p>\n<\/li>\n<li class=\"\" data-start=\"4345\" data-end=\"4386\">\n<p class=\"\" data-start=\"4347\" data-end=\"4386\">Overholdelse af internationale standarder<\/p>\n<\/li>\n<li class=\"\" data-start=\"4387\" data-end=\"4422\">\n<p class=\"\" data-start=\"4389\" data-end=\"4422\">Enkel betjening og dataregistrering<\/p>\n<\/li>\n<li class=\"\" data-start=\"4423\" data-end=\"4476\">\n<p class=\"\" data-start=\"4425\" data-end=\"4476\">Fleksibel tilpasning til specifikke testops\u00e6tninger<\/p>\n<\/li>\n<\/ul>\n<p class=\"\" data-start=\"4478\" data-end=\"4682\">Uanset om du validerer produktdesign eller st\u00e5r for test i forbindelse med frigivelse af produktpartier, <strong data-start=\"4554\" data-end=\"4564\">SLT-01<\/strong>, <strong data-start=\"4566\" data-end=\"4577\">SPPT-01<\/strong>, og <strong data-start=\"4583\" data-end=\"4593\">MST-01<\/strong> levere p\u00e5lidelige og reproducerbare resultater, der er skr\u00e6ddersyet til medicinsk udstyrsbranchens behov.<\/p>\n<hr class=\"\" data-start=\"4684\" data-end=\"4687\" \/>\n<p class=\"\" data-start=\"4704\" data-end=\"5183\">Overholdelse af <a href=\"https:\/\/www.iso.org\/standard\/64790.html\" target=\"_blank\" rel=\"noopener nofollow\"><strong data-start=\"4730\" data-end=\"4744\">ISO 7886-1<\/strong><\/a> er ikke blot en lovgivningsm\u00e6ssig hindring \u2013 det er et tilsagn om kvalitet og patientsikkerhed. Ved at forst\u00e5 og gennemf\u00f8re vigtige unders\u00f8gelser som f.eks. <strong data-start=\"4885\" data-end=\"4912\">test med en injektionsspr\u00f8jte<\/strong>, <strong data-start=\"4914\" data-end=\"4927\">t\u00e6thedspr\u00f8ve<\/strong>, <strong data-start=\"4929\" data-end=\"4956\">test af spr\u00f8jtens d\u00f8drum<\/strong>, og <strong data-start=\"4962\" data-end=\"4982\">glidekraftstest<\/strong>, kan producenterne med tillid bringe deres produkter p\u00e5 markedet. Udstyr dit laboratorium med Cell Instruments\u2019 avancerede testudstyr for at str\u00f8mline testprocessen og sikre n\u00f8jagtighed i alle faser af spr\u00f8jteproduktionen.<\/p>\n<p class=\"\" data-start=\"5185\" data-end=\"5296\"><strong data-start=\"5185\" data-end=\"5296\">Kontakt Cell Instruments i dag, hvis du har brug for teknisk r\u00e5dgivning, hj\u00e6lp til valg af instrumenter eller skr\u00e6ddersyede l\u00f8sninger.<\/strong><\/p>","protected":false},"excerpt":{"rendered":"<p>ISO 7886-1 Hypodermic Syringe Test The ISO 7886-1 standard defines the performance and safety requirements for hypodermic syringes intended for single use, emphasizing aspects like leak resistance, dead space, and glide force. Manufacturers and quality assurance teams must understand the testing procedures outlined in ISO 7886-1 to ensure product compliance and patient safety. ISO 7886-1 &hellip;<\/p>\n<p class=\"read-more\"> <a class=\"\" href=\"https:\/\/www.qualitester.com\/da\/standards\/iso-7886-1\/\"> <span class=\"screen-reader-text\">ISO 7886-1<\/span> L\u00e6s mere \"<\/a><\/p>","protected":false},"author":4,"featured_media":0,"parent":4948,"menu_order":72,"comment_status":"closed","ping_status":"closed","template":"","meta":{"site-sidebar-layout":"default","site-content-layout":"default","ast-global-header-display":"","ast-main-header-display":"","ast-hfb-above-header-display":"","ast-hfb-below-header-display":"","ast-hfb-mobile-header-display":"","site-post-title":"disabled","ast-breadcrumbs-content":"","ast-featured-img":"disabled","footer-sml-layout":"","theme-transparent-header-meta":"default","adv-header-id-meta":"","stick-header-meta":"","header-above-stick-meta":"","header-main-stick-meta":"","header-below-stick-meta":"","footnotes":""},"class_list":["post-6648","page","type-page","status-publish","hentry"],"_links":{"self":[{"href":"https:\/\/www.qualitester.com\/da\/wp-json\/wp\/v2\/pages\/6648","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/www.qualitester.com\/da\/wp-json\/wp\/v2\/pages"}],"about":[{"href":"https:\/\/www.qualitester.com\/da\/wp-json\/wp\/v2\/types\/page"}],"author":[{"embeddable":true,"href":"https:\/\/www.qualitester.com\/da\/wp-json\/wp\/v2\/users\/4"}],"replies":[{"embeddable":true,"href":"https:\/\/www.qualitester.com\/da\/wp-json\/wp\/v2\/comments?post=6648"}],"version-history":[{"count":0,"href":"https:\/\/www.qualitester.com\/da\/wp-json\/wp\/v2\/pages\/6648\/revisions"}],"up":[{"embeddable":true,"href":"https:\/\/www.qualitester.com\/da\/wp-json\/wp\/v2\/pages\/4948"}],"wp:attachment":[{"href":"https:\/\/www.qualitester.com\/da\/wp-json\/wp\/v2\/media?parent=6648"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}